Innovative Technology For Trace Organic Analysis
Seach
TEL: +44 (0)1443 230 935
Mobile Menu

What difference does sorbent mesh size make?

Sorbent mesh size has very little effect on sorbent retention volume, provided that there are at least four or five particles across the internal diameter of the sample tube. This stems from the fundamental theory of gas–solid chromatography, and explains why Carbopack B (60/80 mesh) has the same retention performance as Carbotrap (20/40 mesh) and Carbograph 1TD (40/60 mesh).

Most thermal desorption sorbent tubes use a 100-mesh gauze to support the sorbent, which means that any sorbent with a mesh size or 80 or greater should be supported within the tube. When changing from one mesh size to another, it is important to consider the difference in impedance and ensure that sampling and desorb flows are verified.

UK
UK
Markes International Ltd
Gwaun Elai Medi-Science Campus
Llantrisant
RCT
CF72 8XL, UK
Tel: +44 (0)1443 230935
Germany
Germany
Markes International GmbH
Schleussnerstrasse 42
D-63263 Neu-Isenburg
Frankfurt
Germany
Tel: +49 (0)6102 8825569
USA (West)
USA
Markes International, Inc
2355 Gold Meadow Way
Gold River
California 95670
USA
Tel: 866-483-5684 (toll-free)
Part of Schauenburg International
Top