Industry standard tubes for thermal desorption…and what matters when selecting them

Wednesday, 1 May 2019 at 4:45:PM

For analytical chemists planning to sample VOCs and SVOCs, selection of ‘industry standard’ sorbent tubes is usually an early item on their checklist.

Working with sorbent tubes and caps with a fixed specification is preferred and sometimes mandatory. Uniformly designed and manufactured tubes support reproducibility and can be used across various thermal desorption systems.

Let’s look at what’s understood by ‘industry standard’ tubes and caps and what’s commercially available.

Top 10 things to consider for TD tubes and capsWhat are industry standard sorbent tubes & caps?

Criteria for sorbent tubes and caps have been written into international standard methods for thermal desorption for decades. In 1979, Working Group 5 of the UK Health & Safety Executive specified a 3.5 inch × 0.25 inch o.d. (outside diameter) tube-type diffusive monitor, compatible with thermal desorption, that has since become the ‘industry standard’. See The theory and practice of diffusive monitoring.

Other influential standard methods include:

  • ISO16000-6: Indoor air - Part 6: Determination of volatile organic compounds in indoor and test chamber air by active sampling on Tenax TA sorbent, thermal desorption and gas chromatography using MS or MS-FID specifies tubes with metal screw caps and PTFE ferrules of length 89 mm (3.5 inch).

  • EPA TO-17, Determination of Volatile Organic Compounds in Ambient Air Using Active Sampling onto Sorbent Tubes notes that the most extensively used sorbent tubes are 3½-inch (89 mm) long x ¼-inch (6.35 mm) o.d., stainless steel or glass tubes with a sorbent bed up to 6 cm. It further notes that conditioned or sampled sorbent tubes should always be capped using 0.25 inch brass Swagelok-type screw caps fitted with PTFE ferrules.

  • ISO 16017, Indoor, ambient and workplace air - Sampling and analysis of volatile organic compounds by sorbent tube/thermal desorption/capillary gas chromatography -- Part 1: Pumped sampling, Part 2: Diffusive sampling refers to determination of the minimum air flow rate for thermal desorption using standard 4 mm to 5 mm bore, 6.4 mm o.d. sorbent tubes with the sorbent bed set in the range from 12 mm to 18 mm from the sampling end of the tube.

More methods are detailed in National and international standard methods relating to speciated monitoring of vapour-phase organic chemicals in air.

Markes International play an active role in the continued development of standards in VOC and SVOC analysis and our experts sit on a number of international committees. This helps ensure the company offers thermal desorption users practical insight and advice on consumables compatible with all the established systems.

Thermal desorption tubes

Industry-standard TD tubes can be used in three ways: packed with a single sorbent bed for diffusive or pumped sampling, packed with multiple sorbent beds for pumped sampling or empty for direct desorption. Markes customers can choose from:

  • Universal tubes designed for targeting unknown compounds across a wide range of volatilities

  • Single sorbent tubes: Popularly packed with Tenax, Carbograph 1TD and other sorbents

  • Multi-bed tubes: Commonly used combinations include Tenax/Carbograph 5TD, Carbopack Y/Carbopack B/Carboxen 1003

  • Application-specific tubes: These include tubes packed (and conditioned) in accordance with specific methods like: AirToxics for EPA TO-17, US EPA Method 325 …

  • Empty tubes: are recommended only for advanced users with their own sorbent packing expertise and equipment, or when doing direct desorption of liquids and solids

Caps compatible with industry-standard tubes (0.25 inch/6.35 mm inside diameter) are designed for a number of uses:

  • Long-term storage caps brass or aluminium Swagelok style caps

  • Short-term storage caps also known as push on caps (not for instrument use)

  • Diffusion caps for passive sampling

With specifications for sorbent tubes and caps well established, what else should a buyer look for? Download our infographic on some of the top factors to consider when choosing supplies that complement an industry standard approach.

Plus save 20% on selected products when you buy industry standard tubes & caps.

Nick van Noorden

 


Nick van Noorden joined Markes International in 2018 as Marketing Manager – Supplies. He started his career in scientific publishing and has a background in marketing communications & strategy for organisations in the business, consumer and not-for-profit sectors.

UK
UK
Markes International Ltd
Gwaun Elai Medi-Science Campus
Llantrisant
RCT, CF72 8XL
UK
Tel: +44 (0)1443 230935
Germany
Germany
Markes International GmbH
Bieberer Straße 1-7
63065 Offenbach am Main
Germany
 
Tel: +49 (0)69 6681089-10
USA (West)
USA
Markes International, Inc.
2355 Gold Meadow Way
Gold River, Sacramento
California 95670
USA
Tel: +1 866-483-5684 (toll-free)
China
China
Markes Instruments (Shanghai) Co., Ltd
No. 1 Building, No. 7 Guiqing Road
Xuhui District
Shanghai 200233
P.R. China
Tel: +86 21 5465 1216
Part of Schauenburg International
Top

Hello, how can we help you today?